Use the measure and source functions to test and calibrate almost any process parameter. The Fluke 725Ex Intrinsically Safe Calibrator is a powerful, multi-function calibration solution that offers:
- ATEX II 1 G EEx ia IIB 171°C KEMA 04ATEX 1303X
- I.S. Class I, Division 1 Groups B-D, 171°C compliance
- Measure Volts dc, mA, RTDs, thermocouples, frequency and ohms
- Source or simulate Volts dc, mA, RTDs, thermocouples, frequency and ohms
- Two channel simultaneous source and measure capability for calibration of transmitters
- Power transmitters with internal loop supply
- Store frequently-used test setups for later use
- Pressure measurement to 3,000 psi/200 bar using any of the 8 intrinsically safe Fluke 700PEx Pressure Modules, calibrate pressure transmitters and I/P devices
- Pressure switch test function to capture set, reset and deadband values
- Compact size and weight
- Simple, push-button user interface
- Rugged and reliable, for field use
|
The optional 700PEx Pressure Modules are ATEX-certified Ex II 1 G EEx ia IIC T4. They are also certified by CSA I.S Class 1, Division 1 areas Group A-D per NEC-500.
|
Measurement Accuracy |
Voltage DC |
30.000 V |
0.02%+ 2 counts (upper display) |
10.000 V |
0.02%+ 2 counts (lower display) |
90.00 mV |
0.02%+ 2 counts |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
|
Current DC |
24.000 mA |
0.02%+ 2 counts |
|
Resistance |
0.0 to 400.0 Ω |
0.1Ω (4-wire), 0.15Ω (2- and 3-wire) |
401 to 1500 Ω |
0.5Ω (4-wire), 1Ω (2- and 3-wire) |
1500 to 3200 Ω |
1Ω (4-wire), 1.5Ω (2- and 3-wire) |
|
Frequency |
2.0 to 1000.0 CPM |
0.05 % + 1 count |
1.0 to 1100.0 Hz |
0.05 % + 1 count |
1.00 to 10.00 kHz |
0.05 % + 1 count |
Sensitivity |
1 V peak-to-peak-minimum |
|
Pressure |
Accuracy |
from 0.025% of range using any of 8 intrinsic safe pressure modules. |
|
(for detailed specifications refer to pressure modules in options and accessories) |
|
Modules available for differential, gage, vacuum, absolute, dual and high pressure. |
|
Source Accuracy |
Voltage DC |
100.00 mV |
0.02%+2 counts |
10.000 V |
0.02%+2 counts |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
|
Current DC |
24.000 mA (Source) |
0.02%+ 2 counts |
24.000 mA (Simulate) |
0.02%+ 2 counts |
|
Resistance |
15.0 to 400.0 Ω |
0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
401 to 1500 Ω |
0.5 Ω(excitation current 0.05 to 0.8 mA) |
1500 to 3200 Ω |
1 Ω(excitation current 0.05 to 0.4 mA) |
|
Frequency |
2.0 to 1000.0 CPM |
0.05% |
1.0 to 1100.0 Hz |
0.05% |
1.00 to 10.00 kHz |
0.25% |
Waveform: |
5 V p-p squarewave, -0.1 V offset |
|
RTDs and Thermocouples |
Measure Accuracy: |
NI-120 |
0.2 °C |
PT-100 (385) |
0.33 °C |
PT-100 (393) |
0.3 °C |
PT-100 (JIS) |
0.3 °C |
PT-200 (385) |
0.2 °C |
PT-500 (385) |
0.3 °C |
PT-1000 (385) |
0.2 °C |
Resolution: |
0.1 °C |
J: |
0.7 °C |
K: |
0.8 °C |
T: |
0.8 °C |
E: |
0.7 °C |
R: |
1.8 °C |
S: |
1.5 °C |
B: |
1.4 °C |
L: |
0.7 °C |
U: |
0.75 °C |
N: |
0.9 °C |
Resolution |
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
XK |
0.6°C |
BP |
1.2°C |
|
Source Accuracy: |
NI-120 |
0.2 °C |
PT-100 (385) |
0.33 °C |
PT-100 (393) |
0.3 °C |
PT-100 (JIS) |
0.3 °C |
PT-200 (385) |
0.2 °C |
PT-500 (385) |
0.3 °C |
PT-1000 (385) |
0.2 °C |
Resolution: |
0.1 °C |
Note: |
Accuracy stated for 4-wire measurement. |
J: |
0.7 °C |
K: |
0.8 °C |
T: |
0.8 °C |
E: |
0.7 °C |
R: |
1.4 °C |
S: |
1.5 °C |
B: |
1.4 °C |
L: |
0.7 °C |
U: |
0.75 °C |
N: |
0.9 °C |
Resolution |
J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C |
XK |
0.6°C |
BP |
1.2°C |
|
Specifications |
Ramp Functions |
Source functions: |
Voltage, current, resistance, frequency, temperature |
Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
|
Loop Power Function |
Voltage: |
12 V |
Accuracy: |
10% |
Maximum current: |
22 mA, short circuit protected |
|
Step Functions |
| | | |